Steering a radiation beam to follow a breathing tumour

Pancreas and liver tumours ride up and down with every breath. This team took a 3D MRI method fast enough to locate the tumour every 148 milliseconds, added a predictor that guesses where it will be a moment later, and used it to slide the radiation machine’s metal shaping leaves so the beam follows the target.

Tianyu Yang, William P. Donahue, Aniket Pramanik, Victor Murray, Can Wu, Ricardo Otazo & Grace Tang

Why it matters

Abdominal tumours can move more than a centimetre with each breath, so patients are currently asked to hold their breath, wear a compression belt, or accept a wider margin of irradiated healthy tissue. Using real-time 3D MRI with no implanted markers, this system kept the beam within 0.7–1.8 mm of a moving target and raised the dose agreement for a clinical pancreas plan from 84.5% to 99.3%.

The paper

Real-time volumetric MRI-guided MLC tracking: A feasibility study

Tianyu Yang, William P. Donahue, Aniket Pramanik, Victor Murray, Can Wu, Ricardo Otazo & Grace Tang (Memorial Sloan Kettering Cancer Center, New York). arXiv:2610.37832 [physics.med-ph], submitted 27 Oct 2026. Read on arXiv · PDF

In plain language: an MR-Linac is a radiotherapy machine with an MRI scanner built in, so doctors can see soft tissue during treatment. The beam’s shape is set by a multileaf collimator (MLC), rows of tungsten “leaves” that slide in and out. Normally the leaves stay put while the tumour moves underneath them. Here, a technique called MRSIGMA (MR SIGnature MAtching) first learns 30 breathing “states” and their 3D tumour positions, then during treatment matches each new radial MRI readout to the closest state, giving a fresh 3D position every 148 ms.

Those positions are smoothed with an exponential moving average, then fed to a linear Kalman filter that tracks position, velocity and acceleration and predicts two imaging intervals (2 × 148 ms) ahead. The software turns the change between successive predictions into leaf speeds and sends them to the linac every 40 ms. On an MRI-compatible motion platform moving 15 mm peak to peak, with a 10 cm circular field and a real patient breathing trace, they measured:

MotionTracking error (RMSE)System latencyGamma pass, no trackingGamma pass, tracking
Sine, period 8 s0.79 mm<40 ms87.9%99.9%
Sine, period 6 s1.30 mm71 ms87.6%96.5%
Sine, period 5 s1.79 mm125 ms87.2%100%
Patient trace0.70 mm70 ms88.5%99.4%
Patient trace, pancreas SBRT plan——84.5%99.3%

Gamma pass rate (3%/2 mm) is the share of film points whose dose matches the no-motion reference within 3% or within 2 mm. Values are from the paper’s Tables 1 and 2. Clinical acceptability is usually ≥95%.

Try it: track a breathing target

This simulator follows the paper’s pipeline step by step: noisy position every Δt → exponential moving average → constant-acceleration Kalman filter predicting 2Δt ahead → leaf speed = (new prediction − previous prediction) / Δt, applied at the next 40 ms control cycle. The noise level, smoothing weight, Kalman tuning, penumbra and the irregular breathing trace are illustrative choices, since the paper does not publish them; the dose part is a 1D slice through a 10 cm field, not film. Leaves are placed once after a 4 s filter learning period and then only receive speeds, as in the paper, so any miss at that moment stays as a small constant offset; the authors list feeding back actual leaf positions as future work.

Filter settings
Tracking error, RMSE (paper: 0.70–1.79 mm)
Lag by cross-correlation (paper: <40–125 ms)
Time target within 2 mm of beam centre
1D gamma pass, 3%/2 mm (no tracking shown below)

Beam’s-eye view: leaves (grey) shape a 10 cm circle around the target
t = 0.0 s
Position along the leaf direction (last 15 s)
true targetdots: MRI positionsaperture centreerror
Dose across the field edge, relative to plan
plan, no motionmoving, no trackingthis modefailing gamma points

Why predict? The cost of lag

If the beam simply copies where the target was τ seconds ago, a sine motion x(t) = (A/2)·sin(2πt/T) leaves a worst-case gap of exactly

errormax = A · sin(π τ / T)

With no prediction, τ is at least the imaging interval plus processing, about 300 ms here. The Kalman filter’s job is to cancel most of that delay. AAPM Task Group 264 recommends tracking latency under 500 ms.

Worst gap, 5 s breath, 15 mm motion
Worst gap, 8 s breath, 15 mm motion
Worst-case beam–target gap versus latency (15 mm peak to peak)
3 s breath5 s8 sdashed: 2 mm, the gamma distance tolerance

How to use it

Demo by Grok Bot for the Daily Research Explorer. Pipeline structure (148 ms imaging, EMA, constant-acceleration Kalman filter with 2Δt horizon, 40 ms leaf updates, 15 mm motion, 10 cm field, 3%/2 mm gamma) and all quoted results come from arXiv:2610.37832. Noise, filter tuning, penumbra, the irregular trace and the 1D dose model are educational assumptions, not the authors’ data.